Compliance guide
FDA Enforcement Discretion on Food Traceability Records


Summary
A controlled recall communication identifies affected devices, states the hazard, gives clear product instructions, and provides a response method. A documented tracking process records who was contacted, what each consignee reported, and how nonresponses or incomplete actions were addressed.
Scope and governing framework
FDA states that medical device recalls are usually voluntary and conducted under the policy framework in 21 CFR Part 7. A separate requirement under 21 CFR 806.10 applies to qualifying manufacturer- or importer-initiated corrections and removals. The FDA device recalls page explains both frameworks.
This distinction matters operationally. A Part 806 report to FDA does not replace prompt notification to affected direct accounts. A consignee notice and an FDA report are separate records, with distinct purposes and content.
What a recall communication must convey
Under 21 CFR 7.49, the recalling firm is responsible for promptly notifying each affected direct account. The format, content, and extent of the communication should correspond to the product hazard and recall strategy.
The communication should state:
- The product is subject to a recall or correction.
- Further distribution or use of remaining affected product must stop immediately.
- The direct account should notify downstream customers when appropriate.
- What the recipient must do with the product.
The notice should be brief and specific. Identify the device with enough detail for immediate recognition, such as the brand or common name, size, model or catalog number, UDI or other code, lot number, serial number, and other relevant identifiers. Explain the reason for the recall and the hazard involved, if any. State the specific action required, such as stopping use, segregating inventory, returning, destroying, or correcting the device.
Include a clear response mechanism so the consignee can report whether affected product is on hand and confirm actions taken. A response form can capture receipt, quantities, downstream distribution, product disposition, and the name and date of the person responding. Avoid promotional material and unrelated qualifications that could obscure the recall message.

Communication channel and confirmation
Section 7.49 describes written channels, including conspicuously marked mail communications. For Class I and Class II recalls, and when appropriate for Class III recalls, the letter and envelope should also be marked "urgent." The FDA device page describes "medical device recall" or "correction" as conspicuous marking language.
Telephone or other personal contacts should ordinarily be confirmed by an appropriate written method and/or documented. Record the contact date, person reached, device and lot details discussed, instructions given, and any confirmation sent. A call log alone should not leave the record unclear about whether the recipient received the full written instructions.
No single mailing service is established by § 7.49 as universally required. If postal mailing is appropriate to the recall plan, Certified Mail can provide mailing and delivery records, but those records do not replace the notice copy, recipient response, or follow-up evidence.
Direct-account response tracking
Build a controlled response log from the distribution list used for the recall. Preserve the list version and the date it was approved or extracted. Each consignee record should show the account identity, contact details, affected device identifiers, distribution history, contact method, and response status.
Track the response as a series of dated events, not as one final status field. Record:
- The initial notice date and channel.
- The delivery or contact evidence available.
- Whether the consignee acknowledged receipt.
- The quantity on hand and quantity already distributed, if applicable.
- Downstream notification and action, where required.
- Product disposition or correction status.
- Follow-up attempts and the result of each attempt.
- Any unresolved exception, owner, and next action.
Do not assume a consignee response deadline that is not in the recall instructions or applicable direction. Set response expectations in the communication and recall strategy as appropriate, then track against those stated expectations. Where necessary, § 7.49 calls for follow-up communications to recipients who fail to respond.

Effectiveness checks and exceptions
A response log and effectiveness check serve related but distinct purposes. Under 21 CFR 7.42, the recall strategy specifies the method and level of effectiveness checks, with levels ranging from Level A at 100 percent of consignees contacted to Level E with no effectiveness checks. Checks verify that consignees at the recall depth specified by the strategy received the notification and took appropriate action. The recalling firm ordinarily conducts the checks, and the FDA assists where necessary and appropriate.
For each check, record the consignee, date, method, contact outcome, receipt confirmation, understanding of instructions, and action verified. If the check identifies a failed delivery, unrecognized product, incomplete downstream notification, or unresolved inventory, document the corrective follow-up and its outcome. Preserve the exception history rather than overwriting an earlier "contacted" status.
Evidence-file checklist
Maintain a recall file that allows a reviewer to reconstruct the notice and response process. Include:
- The approved recall strategy and applicable scope.
- The distribution list version used for outreach.
- The final communication and any attachments, with version or approval information.
- A record for each consignee showing channel, date, recipient, and available proof of contact or delivery.
- Consignee replies, including quantities on hand and downstream distribution information.
- Follow-up communications, nonresponse handling, and resolution of exceptions.
- Effectiveness-check plan, sample or account selection, results, and resulting actions.
- Product disposition, correction, or other action records.
For postal communications, retain the mailing record and available tracking or delivery evidence with the final notice. For electronic communications, retain the sent version, transmission information, delivery or failure status, and response records. These materials support a coherent evidence chain from account selection through resolution. Related guidance on proof of mailing versus proof of notice and notice record retention explains broader record-control practices.

Evaluate FDA reporting separately
Under § 806.10, a device manufacturer or importer must submit a report for a correction or removal initiated to reduce a risk to health or to remedy an Act violation that may present a risk to health, subject to applicable exceptions. A required report is due within 10 working days of initiating the correction or removal.
The report includes information such as the device and affected lots, responsible firm, event and action description, quantity and distribution details, consignee names and contact information, and copies of communications. This reporting obligation is separate from sending recall communications to direct accounts. Keep the report, submission evidence, and any amendments in a distinct, cross-referenced section of the recall file.
Operational sequence
A reviewable process follows an orderly sequence: define the recall scope and strategy, verify the affected direct-account list, issue a device-specific communication promptly, capture each account's response, follow up where needed, conduct strategy-based effectiveness checks, resolve exceptions, and separately assess FDA reporting obligations.
This article provides general compliance information, not legal advice.
Frequently asked questions
Summary
A controlled recall communication identifies affected devices, states the hazard, gives clear product instructions, and provides a response method. A documented tracking process records who was contacted, what each consignee reported, and how nonresponses or incomplete actions were addressed.
Scope and governing framework
FDA states that medical device recalls are usually voluntary and conducted under the policy framework in 21 CFR Part 7. A separate requirement under 21 CFR 806.10 applies to qualifying manufacturer- or importer-initiated corrections and removals.
What a recall communication must convey
Under 21 CFR 7.49, the recalling firm is responsible for promptly notifying each affected direct account. The format, content, and extent of the communication should correspond to the product hazard and recall strategy.The communication should state:The product is subject to a recall or correction.Further distribution or use of remaining affected product must stop immediately.The direct account should notify downstream customers when appropriate.What the recipient must do with the product.The notice should be brief and specific.
Communication channel and confirmation
Section 7.49 describes written channels, including conspicuously marked mail communications. For Class I and Class II recalls, and when appropriate for Class III recalls, the letter and envelope should also be marked "urgent." The FDA device page describes "medical device recall" or "correction" as conspicuous marking language.Telephone or other personal contacts should ordinarily be confirmed by an appropriate written method and/or documented.
Direct-account response tracking
Build a controlled response log from the distribution list used for the recall. Preserve the list version and the date it was approved or extracted.